Model | OCMA-500 |
---|---|
Product name | Oil content analyzer |
Measurement method | Solvent extraction - non-dispersive infrared absorption analysis method |
Measured objects | Substances extracted from sample water into solvent and having infrared absorption near a wavelength from 3.4 μm to 3.5 μm |
Measurement range | 0 mg/L to 200 mg/L |
Resolution | For mg/L 0 to 99.9: 0.1, 100 to 200: 1 For mg/g, mg/kg, mg/PC 0 to 9.99: 0.1, 10.0 to 99.9: 0.1, 100 to 200: 1 |
Repeatability | 0 mg/L to 9.9 mg/L: ±0.3 mg/L. 10.0 mg/L to 99.9 mg/L: ±2.1 mg/L. 100 mg/L to 200 mg/L: ±5 mg/L. * For standard liquids |
Display method | 3.5 inches, 320×240 dots, Backlight Color graphic LCD |
Calibration method | Zero, span calibration |
Amount of test sample required | Sample water : Solvent = 2:1 |
Extraction solvent | S-316 |
Amount of extraction solvent required | 8 mL |
Extraction method | Built-in extractor |
Ambient operating temperature | 0°C to 40°C (no condensation) |
Power supply | AC 100 V to 240 V ±10%, 50/60 Hz |
Power consumption | AC 100 V: Approx. 60 VA, AC 240 V: Approx. 90 VA |
External dimensions | 342 (H) × 200 (W) × 313 (D) mm |
Mass | Approx. 7 kg |
External output | Output to a USB memory stick |
Functions |
|