Measurement Range and Repeatability
Unit : mass%
Component | Measurable range | Repeatability | |
---|---|---|---|
1st solution | HNO3 | 30.0 - 34.0 | ±0.5 |
HF | 2.00 - 4.00 | ±0.10 | |
H3PO4 | 35.0 - 39.0 | ±0.6 | |
H2SiF6 | 0.00 - 3.00 | ±0.50 | |
H2O | 25.0 - 33.0 | ±1.0 |
Conditions of measurement
1) Measurement interval: approx. 8 sec.
2) Moving average time :10 times
3) Atmospheric temperature, chemical temperature fluctuation: within /- 1 degree C (period of time: 1hour). Sudden temperature change should be avoided
4) Condition of sample temperature: 20 to 30 degree C
- Repeatability is the maximum deviation from average measured value over a 1 hour continuous measurement period.
- The values in the above table are assigned analog output values between 4 - 20 mA. (makimum resistance of 500Ω)
Specifications Comparison with Previous Model
item | CS-100 | CS-700 |
---|---|---|
Measurement principle | Absorption spectroscopic method | Absorption spectroscopic method |
Sample temperature | 20 to 30 Degree C | 20 to 30 Degree C |
Ambient temperature | 20 to 25 Degree C | 20 to 30 Degree C |
Components | Max. 4 components | Max. 8 components |
Chemical range | Max. 4 ranges | Max. 6 ranges |
Communication | Parallel I/O, Analog output, RS-232C | Parallel I/O, Analog output, RS-232C |
Analog output (4-20 mA) | Max. 2 components | Max. 6 components |
Dimension | 205(W) X 329(D) X 269(H)mm | 218(W) X 391(D) X 213(H)mm |
Power source | DC 24V | DC 24V(Same connecter of CS-100) |
Consumable parts | Halogen lamp | Halogen lamp (Same as CS-100 type) |
Wavelength | UV/NIR | UV/NIR |
Wetted material | PTFE, PFA, Sapphire or Quartz, | PTFE, PFA, Sapphire, Kalrez 4079 O-ring |
Schematics